4
1/2 DIGIT THERMOCOUPLE METERS
MODEL 2002B-J, -K, -T / 3002-J, -K, -T
|
Models 2002B-J, -K and -T
are low-cost 4 1/2 digit (±19,999 count) thermocouple meters
with 0.1°C or 0.1°F resolution for indication-only applications.
High accuracy is provided by five-segment linearization and matched,
graded components for each T/C type.
STANDARD FEATURES
Mechanical features include a bright 0.56 in (14.2 mm) display,
a screw-terminal barrier strip for signal and power, and a compact
1/8 DIN case that requires less than 4.1 in (104 mm) behind the
panel. Electrical features include cold-junction compensation, jumper
selection of °C or °F, and a linearized recorder output
of 1 mV per degree.
OPTIONS
Power options include isolated 9-32 Vdc and 26-56 V dc. A splash-proof
lens cover which meets NEMA-4 requirements allows the 2002B to be
used in harsh industrial environments.
SPECIFICATIONS
INPUT
Thermocouple:
Type J (Iron-Constantan, Fe-CuNi)
Type K (Chromel-Alumel, NiCr-NiAl)
Type T (Copper-Constantan, Cu-CuNi)
Calibration: IEC 584-1 (IPTS-68)
Configuration: Single-ended (-TC lead tied to ANA GND)
Polarity: Bipolar
Zero: Adjustable ±20°C (±36°F)
Cold-junction tempco: ±0.05 deg/deg
Sensor-wire resistance effect per conductor:
Type J: 40 µdeg/deg/ohm, up to 2500 ohm
Type K: 50 µdeg/deg/ohm, up to 2000 ohm
Type T: 50 µdeg/deg/ohm, up to 200 ohm
Sensor-break current: 300 nA
Sensor-break indication: Positive overrange (upscale)
NOISE
REJECTION
NMR, SIG HI to SIG LO: 70 dB, 50/60 Hz
CMR, ANA GND to PWR GND: 120 dB, DC to 60 Hz
CMV, ANA GND to PWR GND: 1500 Vp per HV test, 354 Vp per
IEC spacing
ACCURACY
AT 25°C
Range
for rated accuracy:
Type J: 138.0 to +760.0°C (216.0 to +1400.0°F)
Type K: 105.0 to +1248.0°C (157.0 to +1999.9°F)
Type T: 112.0 to +400.0 °C (170.0 to +752.0°F)
Maximum error: ±1.5°C (±2.7°F) ±5
counts
Span tempco: ±0.01% of reading/°C
Resolution: 0.1°C or 0.1°F (jumper-selectable)
Full-scale step response: 1 s
Warmup to rated accuracy: 20 min
ANALOG OUTPUT
Voltage:
1 mV/deg
Calibration error: ±1.5 mV on °C, ±2.7 mV
on °F
Current: 1 mA max
Source resistance: 68 ohm
ANALOG-TO-DIGITAL
CONVERSION
Technique: Dual-slope, average value
Signal-integration period: 100 ms
Read rate: 2.5/s
DISPLAY
Type: 7-segment, red LED
Height: 0.56 in (14.2 mm)
Symbols: 1.8.8.8.8
Decimal points: Four positions programmable by jumpers behind
the lens or at connector
Overrange indication: Four least-significant digits flash
DIGITAL
INPUTS
Level:
TTL or 5 V CMOS compatible
POWER
AC voltages: 115 or 230 Vac, ±15%
AC frequency: 49-63 Hz
DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated
to 300 Vp
Power consumption: 3.7 W nominal
Output voltages: +4.7 Vdc and 4.7 Vdc ±5%, 10
mA max
ENVIRONMENTAL
Operating temperature: 0 to 60°C
Storage temperature: -40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)
MECHANICAL
Dimensions: Newport DIN2A (104 mm 1/8 DIN case) (see Mechanical
section for drawings)
Weight: 15 oz (425 g)
Case material: 94V-0 UL-rated polycarbonate
|
|
|
† All dollar amounts on this site are shown in US currency.
|
|
 |
 |
 |
 |
 |
 |
 |
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
Model
|
| |
Select 2002B for Standard Meter 3002 for Meter with INFINITY® Style Bezel and Adapter
|
(2)
|
INPUT CONFIGURATION
|
| |
Select -JC1 for -138 to +760°C. (JB1) -JF1 for -216 to +1400°F (JB1) -KC1 for -105 to +1248°C (KB1) -KF1 for -157 to +1999°F (KB1) -TC1 for -112 to +400°C (TB1) -TF1 for -170 to +752°F (TB1) -MC for -200 to +820°C (MB) -MF for -328 to 1508°F (MB) -OR1 for 19.99 ohms full-scale (RB) -OR2 for 199.9 ohms full-scale (RB) -OR3 for 1999.9 ohms full-scale (RB) -OR4 for 199.99 kOhms full-scale (RB)
|
(3)
|
POWER
|
| |
Select Nothing(leave field blank) for 115 Vac, 50/60 Hz ,C1 for 230 Vac, 50/60 Hz ,C3C for Isolated 9-32 Vdc ,C3E for Isolated 26-56 Vdc
|
(4)
|
DISPLAY OPTION
|
| |
Select Nothing(leave field blank) for Red LED display ,G for Green LEDs for Display
|
(5)
|
ADD-ON OPTIONS
|
| |
Select Nothing(leave field blank) for No Options ,BL for Lens without Newport logo
|
| NOTE: All combinations may not be valid, check spec sheet for valid part numbers. |
| | | |
|
 |
 |
|
|
|
|
Keywords:4 1/2 DIGIT THERMOCOUPLE METERS
|
Back to top |
|
|