4 1/2 DIGIT STRAIN METERS WITH BCD OUTPUT
MODEL 2003B-LA1 & 2003B-LP1
OPTION
AVG/RMS
|
Models 2003B-LA1
and 2003B-LP1
are high-resolution, high-accuracy 4 1/2 digit panel meters for use
with pressure transducers, load cells and other bridge transducers
which require excitation and preamplification of microvolt signals.
Model 2003B-LA1 provides an extra 2 Hz two-pole active filter
for exceptional normal mode noise rejection (NMR) of 110 dB.
Model 2003B-LP1 provides an analog peak-hold circuit with external
reset. It is ideal for destructive testing applications.
EXCITATION SUPPLY AND PREAMPLIFIER
The 2003B-LA1 and 2003B-LP1 provide an isolated, constant-voltage
excitation supply which can be field-configured for 5, 10 or 15 V
dc output. They also provide a high-impedance, true-differential preamplifier,
which is continuously adjustable from 20 to 100 in 3% steps to accommodate
load-cell sensitivities of 1 mV/V to 20 mV/V. Fine-gain and fine-deadload
adjustments are easily accessible behind the lens. A push-to-cal switch
is provided through the lens. The calibration shunt resistor is customer-supplied.
BRIDGE CONNECTION
Bridges can be connected by 4 or 6 wires. In 6-wire connection,
two lines sense the voltage applied to the bridge and compensate for
lead resistance.
RESOLUTION AND ACCURACY
Resolution is one part in 19,999 counts, or 0.005% of full scale.
Accuracy is up to 99.98%. of reading. The measurement is ratiometric
to reduce errors due to temperature and power-line variations. The
same internal reference controls the meter gain and the excitation
level.
|
SIGNAL
OUTPUTS
Standard outputs include non-isolated parallel BCD and
a strip-chart recorder output of 1 mV/count (nominal).
OPTIONS
Options include board-edge connectors and a splash-proof
lens cover which meets NEMA-4 requirements. For additional
information, please refer to the Options and Mechanical sections.
|
|
SPECIFICATIONS
|
ANALOG INPUT
|
Most-sensitive
scaling
|
Least-sensitive
scaling
|
|
Signal
Range
|
0.5
to +2 mV
|
2.5
to +10 mV
|
|
Display
Range
|
5000
to +19,999
|
5000
to +19,999
|
|
Resolution
|
1
µV/count
|
5
µV/count
|
|
Preamplifier
Gain
|
100
|
20
|
Input configuration:
Differential, ratiometric
Coarse-span adjustment: 57 overlapping ranges with fine adjustment
Fine-span adjustment: ±3% of span
Zero adjustment: ±3% of bridge output
Deadload adjustment, max: Up to 60% of bridge output with external
resistor
Tare adjustment: ±4 counts nominal
Bias current: 11 nA typ, 22 nA max
EXCITATION
Configuration:
Constant-voltage
Voltage sense: Internal or remote
Voltage settings: Jumper-selectable 5, 10, 15 Vdc
Load current, max: 125 mA at 5 V, 115 mA at 10 V, 67 mA at
15 V
NOISE
REJECTION
NMR, SIG HI to SIG LO, LA1 option: 110 dB, 50/60 Hz
NMR, SIG HI to SIG LO, LP1 option: 80 dB, 50/60 Hz
CMR, SIG GND to SIG LO: 80 dB, DC to 60 Hz
CMR, SIG GND to PWR GND: 120 dB, 50/60 Hz
CMV, SIG GND to PWR GND: 1500 Vp test, 354 Vp per IEC spacing
ACCURACY AT 25°C
Error, max: ±0.01% of reading ±2 counts
Zero tempco: ±0.3 µV/°C
Span tempco: ±0.005% of reading/°C
Step response: 1 s to 99.9% of span
Warmup to rated accuracy: 60 min
CONVERSION
Technique: Dual-slope, average-value
Signal integration period: 33 ms at 60 Hz, 40 ms at 50 Hz
Read rate: 4.3/s at 60 Hz, 3.6/s at 50 Hz
DISPLAY
Type: 7-segment, red LED
Height: 0.56 in (14.2 mm)
Symbols: 1.8.8.8.8
Decimal points: Four positions programmable internally or at
connector
Overrange indication: Display flashes
DIGITAL SIGNALS
Output signals: TTL-compatible
HOLD input: TTL or 5 V CMOS compatible
BLANKING input: Open-collector compatible
POWER
AC voltages: 115 Vac ±10% at 60 Hz; 230 Vac ±10% at
50 Hz
Power consumption: 8 W (nominal)
ENVIRONMENTAL
Operating temperature: 0 to 50°C
Storage temperature: 40 to +75°C
Relative humidity: 95% at 40°C (non-condensing)
MECHANICAL
Dimensions: Newport DIN1A (1/8 DIN) case (see Mechanical
section for drawings)
Weight: 23 oz (650 g)
Case material: 94V-0 UL-rated polycarbonate
|
|
|