Low cost,
small size, low power consumption at 5 Vdc and exceptional power-supply noise
rejection make the 2520-PA ideal for demanding OEM and
portable-instrument applications.
MICROVOLT PREAMPLIFIER The 2520-PA
incorporates a preamplifier, which can be programmed for resolution from 1 to 25
µV/count and for zero offset up to ±0.6 mV (RTI) ±21% of full-scale readout
(RTO). The 2520-PA can be configured for use with load cells and pressure
gauges, which require zero and span adjustment for readout in engineering units.
Bridge excitation may be provided by the same 5 Vdc supply that powers the
meter, or by an external, isolated 10 or 15 Vdc power supply. The 2520-PA can
also be configured as a microvoltmeter using an internal reference, or in a
ratiometric mode using an external reference.
|
SERIAL BCD Non-isolated,
character-serial BCD output is standard to allow the 2520-PA to be
interfaced to other instrumentation.
Model 2520-PA provides
a differential programmable preamplifier for use with low-level bridge
circuits. Bridge excitation can be provided by the same 5 Vdc power supply
that powers the meter. |
|
SPECIFICATIONS
ANALOG
INPUT Signal
range: ±20 mV (Most Sens. Scaling) / ±500 mV (Least Sens.
Scaling) Display range: ±19,999 (Most Sens. Scaling) / ±19,999 (Least
Sens. Scaling) Resolution (µV/count): 1 (Most Sens. Scaling) / 25
(Least Sens. Scaling) Preamplifier gain: 100 (Most Sens. Scaling) / 4
(Least Sens. Scaling) Input Configuration: Differential,
ratiometric Coarse-Span Adjustment: Jumper-selectable gain
ranges Fine-Span Adjustment: Front-panel accessible precision
potentiometer Zero Adjustment (RTI or Deadload): ±0.6 mV Zero
Adjustment (RTO or Tare): ±20.9% of full-scale display Bias
Current: 10 nA typ, 20 nA max
NOISE
REJECTION NMR, SIG HI to SIG LO: 60 dB, 50/60 Hz CMR,
SIG GND to SIG LO: 86 dB, dc to 60 Hz CMV, SIG GND to SIG LO: -1
to +2.5 Vdc Power Supply Rejection: 86 dB at 50/60 Hz
ACCURACY AT 25°C Error Max: ±0.01% of
reading ±2 counts Span Tempco, Typ: ±0.003% of reading/°C Zero
Tempco, Typ: ±0.3 µV/°C Step Response: 2 sec to 99.9% of
span Warmup to Rated Accuracy: 30 min
CONVERSION Technique: Dual-slope,
average-value Signal Integration Period: 100 msec Read Rate:
2.5/sec
DIGITAL SIGNALS Output
Signals: TTL-compatible HOLD Input: TTL or 5 V CMOS
compatible BLANKING Input: Open-collector compatible
POWER Voltage: +5 to +5.25
Vdc Current: 170 to 230 mA Power Consumption: 1 W
(nominal)
ENVIRONMENTAL Operating
Temperature: 0 to 55°C (32 to 122°F) Storage Temperature: -40 to
+85°C (-40 to 184°F) Relative Humidity: 95% at 40°C (104°F)
(non-condensing)
MECHANICAL Dimensions: 24 H x 96 W x 72 mm
D (0.94" x 3.77" x 2.82") Panel Cutout: 22.2 x 92 mm Weight:
145 g (5 oz) Case Material: 94 V-0 UL-rated
polycarbonate
|
|
|
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with operator's manual.
Ordering Example: (1) 2520-PA Strain/Microvoltmeter with heavy duty U-clamp mount = $370
|
|
|
|
|
|
|