Linear
Thermistor Components and Probes & Component Applications
44000 Series
 |
Linear
Response Components For applications requiring thermistors with
linear response to temperature change, NEWPORT offers linear
components. These unique devices consist of a thermistor composite
for temperature sensing and an external resistor composite for
linearizing.
Thermistor composites 44018 and 44019 each contain two thermistors
packaged in a single sensor (Figures 1A & 1B). Thermistor composite
44020 contains three thermistors packaged in a single sensor
(Figure 1C). |
Resistor composites for use with 44018 and 44019 thermistor composites
consist of two metal film resistors of the size shown in Figure 2.
Resistor composites for use with the 44020 thermistor composite consist
of three of the same type metal film resistors.
Linear components are manufactured with different values for different
temperatures ranges. When they are connected in networks shown in
Figures 3 and 4, they produce a varying voltage or resistance which
is linear with temperature.
One of the basic network manifestations is a voltage divider as in
Figure 3A for components other than #44212, and as shown in Figure
3B for component #44212. The area within the dashed lines represents
the thermistor composite. The network hookup for linear resistance
versus temperature is shown in Figure 4A for linear components except
#44212, and in Figure 4B for #44212.

Following is a
description of why these networks produce linear information. The
equation for a voltage divider network, consisting of R and R0
in series, is:
where Eout
is the voltage drop across R. If R is a thermistor, and Eout
is plotted versus temperature, the total curve will be essentially
non-linear and of a general “S” shape, with linear or nearly linear
portions near the ends and in the center.
If R is modified by the addition of other thermistors and resistors,
linearity of the center section of the curve, where sensitivity is
greatest, can be extended to cover a wide range of temperatures. This
section follows the general equation for a straight line, y = mx +
b or in terms of a linear component:
|
For
Voltage Mode
|
For
Resistance Mode
|
|
Eout
= ±MT + b
|
Rt
= MT + b
|
|
where
M is slope in volts/°T,
T is temperature in °C or °F,
and b is the value of Eout when T = 0°
|
where
M is slope in ohms/°T,
T is temperature and in °C or °F,
and b is the value of the total network resistance,
Rt, in ohms when T = 0°
|
Linear
Voltage vs. Temperature
Figure
3A

|
Figure
3B
|
Note: Model
5830 precision benchtop thermometer includes linearized circuity.
Linear
Resistance vs. Temperature
Figure
4A

|
Figure
4B

|
*RL1
may be any value as long as a new R1 value
(R1A) is selected to satisfy the relationship:
|
R1A
=
|
R1
x RL1
|
|
RL1
– R1
|
Sensitivity is 400 times greater than an IC thermocouple. Thermistor
values as high as 30 mV/°C are common. In addition, output voltage
can be applied to a recorder or digital voltmeter to produce a precise,
sensitive, direct reading thermometer.
Multiplexing
The 44018 thermistor composite is used in four of the linear components.
The part that changes in each component is the resistor composite,
which determines the temperature range. Therefore, the 44018 thermistor
composite can be used over the entire -30 to 100°C temperature range
by simply changing resistor composites. Its accuracy and interchangeability
over the full range is ±0.15°C. It is not mandatory that NEWPORT®
resistor composites be used with the 44018 thermistor composite. Any
0.1% resistors of the proper values and with a temperature coefficient
of 30 PPM or less may be substituted. In other situations, it is frequently
desirable to have thermistor composite temperature sensors at more
than one location. When this is required, it is not necessary to have
a separate resistor composite for each thermistor composite. It is
possible to multiplex any number of thermistor composites through
a single resistor composite for greater design flexibility: Linear
Thermistor Components are manufactured under U.S. Patent #3316765
and Canadian Patent #782790.
To Order (Specify Model Number)
|
Linear
Kit† Mdl. #
|
Thermistor
Composite Mdl. #
|
Resistor
Composite Mdl. #
|
|
44201
|
44018
|
44301
|
|
44202
|
44018
|
44302
|
|
44203
|
44018
|
44203
|
|
44204
|
44018
|
44304
|
|
44211A
|
44019
|
44311A
|
|
44212
|
44020
|
44312
|
†Kit
includes thermistor composite and resistors.

Component
Specifications

|
*
Ein Max. and
*IT Max values have been assigned
to control thermistor self-heating errors so they
do not enlarge the component error band; i.e., the
sum of the linearity deviation plus the probe tolerances.
The values were assigned using a thermistor dissipation
constant of 8MW/°C in stirred oil. If better heat-sink
methods are used or if an enlargement of the error
band is acceptable, Ein Max. and IT Max
values may be exceeded without damage to the thermistor
probe.
***See Figure 1, example 1 on typical linear
component application page.
†Kit includes thermistor composite and resistors.
|
|
**
The maximum error at any point is the algebraic sum of the
thermistor manufacturing tolerances, plus linearity deviation,
a fixed network behavior. Condition "A" is the worst case
linearity deviation of ±0.15°C and may occur with the ±0.1%
resistors supplied. Condition "B" exists when the three
resistors are whin ±0.02% of nominal, which reduces linearity
deviation to ±0.08°C.
Note: The time required for a thermistor composite
to indicate 63% of a newly impressed temperature is one
second in "well stirred" oil and ten seconds in free, still
air.
|
|
|
Typical
Linear Component Applications
Example 1:
To measure and record on a 100 mV recorder temperature in the
range 30 to 40°C.
1. Select Part number 44202 (temperature range -5°to +45°C)
basic equation Eout1 = (-0.0056846 Ein) T +0.805858
Ein
2. Calculate Ein for 10°C equal to 100 mV
(Eout, @30°C - Eout1 @ 40°C) = 100 mV
[(-0.0056846 Ein) 30°C + 0.805858 Ein] - [(-0.0056846
Ein) 40°C + 0.805858 Ein] = 100 mV
0.056846 Ein = 100 mV
Ein = 1.7591 Volts
3. Using the Linear network as two legs of a Wheatstone bridge
add the two additional legs, R3 and R4 so that
Eout2 = 0 when T = 30°C. (See Figure 1.) R3
and R4 are calculated from five known conditions.
(1) The voltage drop across R4 (ER4) should
equal Eout1 at 30°C for Eout2 to equal zero.
(2) Ein = 1.7591 Volts
(3) 1000 ohms < or equal to R3 + R4 <
or equal to 5000 ohms. (If R3 + R4 is less than
1 K, excessive battery drain may occur. If R3 + R4
is more than 5 K, some degradation of linearity will occur.)
(5) Eout1
= -0.0056846 (1.7591 Volts) (+30°C) +0.805858 (1.7591 Volts) = 1.1180
Volts
| ER4
= Eout1 = ER4 = |
Ein
R4
R3 + R4
|
or
1.1180 =
|
R4
1.7591
R3
+ R4
|
and
let us choose R3 + R4 = 1000 ohms. |
| Solve
for R3 and R4 |
1.1180
=
|
R4
1.7591
R4
+ 100-R4
|
R4
= 635.55 ohms
R3 = 364.45 ohms |
4. Apply Eout2 to the recorder input terminals and the
result is a direct reading 10°C full scale thermometer.
Example 2: To
make a 4 digit 100 mV sensitivity digital voltmeter into a direct
reading differential thermometer whose ambient range is -30 to 40°C;
1. Select Part number 44203 (temperature range -30 to 50°C)
basic equation Eout = (-0.0067966 Ein) T +0.65107 Ein
2. Calculate Ein so that 10 mV equals one degree C. (This is
done so that the Digital Volt Meter will read directly in temperature
with 0.01°C readability)
(Eout, @ -30°C - Eout, @ +40°C) = 0.700 Volts
[(-0.0067966 Ein)(-30) +0.65107 Ein] - [(-0.0067966
Ein) (40) + 0.65107 Ein] = 0.700
0.47576 Ein = 0.700
Ein = 1.4713 Volts
3. Connect two linear networks (#44203) as shown in Fig. 2.
4. Apply Eout to the Digital Volt Meter input terminals
for a direct reading differential thermometer.

Example 3:
To make a 2-wire system from a 3-wire system using any Linear component:
1. For voltage mode, connect R2 to the thermistor
composite. (See Figure 3.) This unit can function as the temperature
sensor and be located remote from the signal conditioning circuit
by up to distance "D".
2. The resistance mode differs from the voltage mode only by
removal of the power source. (See Figure 4.)
3. Acceptable distance "D" varies according to the temperature
range. Using #22 wire "D" may be as follows without loss of accuracy
in both 2-wire and 3-wire systems. Where distance "D" is greater than
indicated, heavier gauge wire may be used.
|
Temperature
Distance Range
|
Distance
Range “D”
|
|
0
to 100°C
|
100 ft.
|
|
-5
to 45°C
|
300
ft.
|
|
-30
to 50°C
|
300
ft.
|
|
+30
to 100°C
|
300
ft.
|
Example 4: Multiplexing to connect any number of thermistor
composites to a single signal conditioning circuit. (See Figure 5.)
Multiplexing can be accomplished much more easily with a two-wire
system, such as shown in Figure 5.
Lead Colors:
Green: Common to T1 & T2
Brown: T1
Red: T2
|
|
|
† All dollar amounts on this site are shown in US currency.
Ordering Example: (1) 44204 Linear thermistor kit with -2 to 38°C (30 to 100°F) temperature range = $37
|
|
|
|
|
|
|
Keywords:Linear Thermistor Components, Probes, and Applications
|
Back to top |
|
|