Virtual Chart Recorder
View Temperature + Humidity with a Web Browser
| Features |
 |
Virtual Chart Recorder |
 |
Web Server |
 |
Alarms by Email or Text Message |
 |
Data Logging |
 |
Accurate Readings |
 |
iTHX-SD: Record Years of Data on Popular SD Cards |
 |
No Special Software Required |
|
|
|
| |
The NEWPORT® iTHX transmitter let's you monitor and record Temperature, Relative Humidity and Dew Point over an Ethernet network or the Internet with no special software except a Web Browser.
The iTHX serves Active Web Pages to display real time readings, display charts of temperature and humidity, or log data in standard data formats for use in a spreadsheet or data acquisition program such as Excel or Visual Basic.
The virtual chart viewed on the web page is a JAVA™ Applet that records a chart over the LAN or Internet in real time.
With the iTHX, there is no need to invest time and money learning a proprietary software program to log or chart the data.
Adjustable Charts
Chart scales are fully adjustable on the fly. For example, the chart can display one minute, one hour, one day, one week, one month or one year. Temperature and humidity can be charted across the full span (-40 to 124°C, and 0 to 100% RH) or within any narrow range such as (20 to 30°C). When a second sensor is added, users can select a chart that records channel 1, channel 2 or the difference of the two channels.
Display and Chart Two Channels
The iTHX transmitters come complete with a temperature and humidity probe for measurement of a single location.
With the addition of a second probe, the iTHX transmitter can measure and display humidity, temperature and dew point in a second location up to ten feet away.
The transmitter can display and chart absolute measurements in both locations, or a differential measurement between the two locations.
The second probe requires no change to the basic iTHX transmitter hardware.

A second probe can be added at the time of purchase or in the future. NEWPORT offers a choice of industrial probes in 2" and 5" lengths, and a wand style for ambient indoor applications.
A simple DB-9 "Y" connector is available for adding a second probe to model iTHX-W.
No connector is required to add a second probe to the DIN rail mounted iTHX-2.
Award-winning Technology
The NEWPORT iTHX is simple to install and use, and features NEWPORT's award-winning iServer technology that requires no special software except a Web Browser. The iTHX connects to an Ethernet Network with a standard RJ45 connector and sends data in standard TCP/IP packets. It is easily configured with a simple menu using a Web Browser and can be password protected. From within an Ethernet LAN or over the Internet, the user simply types its IP address or an easy to remember name such as "Cleanroom5" or "ServerRoom"
in any Web Browser, and the iTHX serves a Web Page with the current readings.
Email Alarms
All iTHX models that are on a LAN that is connected to the Internet can trigger an alarm that can be sent by email to a user or a distribution list anywhere in the world, including text messages to cell phones and PDA's.
Typical Applications
The iTHX is great for monitoring temperature + humidity in applications such as: clean rooms, computer rooms, HVAC systems, pharmaceutical and food processing and storage, hospitals, laboratories, semiconductor fabs, electronic assembly, warehousing, museums, manufacturing, greenhouses, farm animal shelters, and many more.
NEW iTHX-SD with SD Flash Memory and LCD Display
Click here for complete specification and ordering
details.
The NEWPORT® iTHX-SD adds several valuable features in addition to the
backlit local display of
temperature and humidity.
2 GB SD Flash Memory Card The iTHX-SD comes complete with a removable 2 GB SD Flash Memory card that
can store up
to seven years of readings taken at ten second intervals.
Records on SD Flash Cards The data recorded can be read on a PC or MAC with a USB card reader or
downloaded
remotely over an Ethernet network or the Internet.
Alarm Relays Two 1.5 Amp relays can be programmed for any combination of temperature or humidity, and
high or low set points.
Battery Backup A standard 9 Volt Alkaline battery (included) allows the device to log data for up to
2 days without external ac power.
Link to Web CAM or IP Camera If you get a message about an alarm condition, you can quickly click on the link to view the actual scene over the Internet.
NEWPORT iServer products are designed and manufactured in Santa Ana, California.
Specifications
Sensor Specifications
|
Relative Humidity (RH)
Accuracy/Range: ±2% for 10 to 90%; ±3% for 5 to 10% and 90 to 95%; ±4%
for 0 to 5% and 95 to 100%
Non-linearity: ±3%
Hysteresis: ±1% RH
Response Time: 8 seconds, tau 63%
Repeatability: ±0.1%
Resolution: 0.1%, 12 bit
|
|
Temperature(T)
Accuracy/Range*:
Wand Probe:
±0.5°C for 5°C to 45°C ( ±1°F for 41°F to 113°F);
±0.5° to ±1°C for 0° to 5°C and 45° to 70°C (±1° to ±2°F for 32° to 41°F and 113° to 158°F)
Industrial Probe:
±0.5°C for 5° to 45°C (±1°F for 41 to 113°F);
±0.5° to ±1.5°C for -40° to 5°C and 45° to 124°C (±1° to ±2.7°F for -40° to 41°F and 113° to
255°F)
(see chart in manual)
*Note: extended temperature range is for Industrial probe only, the iServer's operating temp. is 0 to
70 °C.
Response Time: 5 to 30 seconds, tau 63%
Repeatability: ±0.1°C
Resolution: 0.1°C, 14 bit
|
Probe Dimensions
Wand Probe iTHP-W-6:
19.1mm Diameter x 198mm Long (0.75" x 7.8")
Cable with DB9 connector: 152mm Long (6")
Cable operating temperature: 0 to 80°C (32 to 176°F)
Industrial Probe iTHP-2, iTHP-5:
16mm Diameter x 51mm Long (0.63" x 2") or
16mm Diameter x 137mm Long (0.63" x 5")
Cable with DB9 connector or stripped wire leads: 0.9m Long (3') or
Cable with DB9 connector or stripped wire leads: 3m Long (10')
Cable operating temperature: -40 to 125°C (-40 to 257°F)
Optional Caps available for Industrial Probes
iServer Specifications
Interfaces
Ethernet (RJ45): 10BASE-T
Sensor: Digital 4-wire (DB-9) for iTHX-W;
removable 8 posistion screw terminals for iTHX-2
Supported Protocols
TCP, UDP, ARP, ICMP, DHCP, DNS, HTTP, and Telnet
Management
Device configuration and monitoring through embedded WEB server (Fig.1)
Embedded WEB Server
Serves WEB pages containing real-time data (Fig. 2)
and live updated charts within definable time intervals (Fig. 3).

Power
Power Input for iTHX-W: 9 to 12 Vdc
Consumption: 2.5 W max.
Safety Qualified ac power adapter:
Nominal Output: 9 Vdc @ 0.5A included
Input: 100 to 240 Vac, 50/60 Hz
Power Input for iTHX-2: 10 to 32 Vdc
Consumption: 2.5 W max.
Switching Power Supply (iDRN-PS-1000)
sold separately for iTHX-2

Environmental
Operating Temperature: 0 to 70°C (32 to 158°F)
Storage Temperature: -40 to 125°C (-40 to 257°F)
Power Adapter Operating Temperature: 0º to 40ºC (32º to 104ºF) for iTHX-W
Packaging
Material: Valox 364 PBT case with wall mount bracket for iTHX-W;
Polycarbonate case with DIN Rail mount for iTHX-2
Dimensions: 27.4H x 66W x 93D mm (1.08 x 2.60 x 3.67 in) iTHX-W
90.2H x 25.1W x 115.0Dmm (3.54 x 0.99 x 4.53 in) iTHX-2
Refer to Mechanical Section
iTHX-W
iTHX-2
Weight: 0.18 kg (0.4 lbs.) iTHX-W
0.113 kg (0.25 lbs.) iTHX-2
Agency Approvals
|
|
|
|
|
† All dollar amounts on this site are shown in US currency.
|
|
 |
 |
 |
 |
 |
 |
 |
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
iServer MicroServer™ Models
|
| |
Select iTHX-W for iServer MicroServer™ for Temperature + Humidity + Dew Point iTHX-2 for DIN Rail mount iServer MicroServer™ for Temperature + Humidity + Dew Point
|
(2)
|
Probe Options
|
| |
Select
-2 for Industrial 2" Probe and 3' Cable with (substitution for standard Probe) -5 for Industrial 5" Probe and 10' Cable (substitution for standard Probe-not available with iTHX-2)
|
| NOTE: All combinations may not be valid, check spec sheet for valid part numbers. |
|
|
 |
 |
|
|
|
|