Navigation image map
Contact Us Cart Site Map Manuals Order Products Home

Frequency Meter, Tachometer & Rate Meter
MODEL Q2000H (3 1/2 digits) Q9000H (4-digits)
Replacement Model - INFCTRA

Electromagnetic Compatibility


NEWPORT PRODUCT INFO
MANUAL - PDF Version Qxxxx-H
MECHANICAL 1/8 DIN (1A) Case
MECHANICAL Connections Diagram
PRICE
REQUIRES ADOBE ACROBAT - HELP
QUANTA STANDARD FEATURES
1,999-count display (Q2000H)
9,999-count display (Q9000H)
Input of 100 Hz to 20 kHz for full-scale display
Front-panel accessible adjustments for zero, span and trigger level
Recorder output Bright, 0.56 in (14.2 mm) LED display
Display hold and test
120/240 Vac, 5 Vdc, 9-32 Vdc or 26-56 Vdc power
Screw-terminal barrier strip
1/8 DIN case
   
OPTIONS
Analog output for user-selected span
Single-setpoint 10 A relay
Dual-setpoint 10 A relays
Proportional 4-20 mA control
Time-proportional control
Parallel BCD output, isolated
Liquid crystal display (Q2000H)
NEMA-4 splash-proof lens cover

The Q2000H and Q9000H
are frequency meter/tachometers with extensive zero and scale factor adjustment capability for direct readout in engineering units for RPM, flow and speed. Input frequencies can be as low as 100 Hz or as high as 20 kHz for the display maximum, which is 1,999 for the Q2000H and 9,999 for the Q9000H. Zero offset is available for use with transducers with a frequency output and a live zero.

EASY TO INTERFACE
The Q2000H and Q9000H are designed for easy interfacing to transducers with a frequency or pulse output. These include turbine flowmeters, passive magnetic pickups, and active proximity detectors with NPN, PNP or NAMUR (1 mA/3 mA) output. To accommodate different signal levels and provide noise immunity, the DC trigger level can be adjusted from ­1 to +5 V, and two hysteresis levels are selectable. Signals as low as 15 mV p-p or as high as 130 V rms can be monitored directly.


CONFIGURATION AND CALIBRATION
The price of the Q2000H or Q9000H includes configuration and calibration, normally provided by an authorized Newport distributor. The units can also be configured and scaled in the field. Adjustments for zero, span, and trigger level are accessible from the front panel to allow final calibration or recalibration without removal of the meter from the panel.

POWER AND DISPLAY OPTIONS
Six types of meter power are available: 120 Vac, 240 Vac, 24 Vac, 5 Vdc, isolated 9-32 Vdc and isolated 26-56 Vdc. An LED display is standard. An LCD display is available with the Q2000H and is ideal for viewing in bright ambient light.

Q2000H orQ9000H scaled to read RPM
The QUANTA Q2000H OR Q9000H Can be scaled to read RPM directly. With the dual-setpoint controller(optional), they can provide under and overspeed alarms. The 1 mV/count analog output (standard) can be used to drive a strip-chart recorder for permanent.

ANALOG OUTPUT OPTIONS

An analog output which tracks the display is standard and is useful for driving a strip-chart recorder. An additional analog output can be provided by an optional vertical plug-in board. Available output signals are 0-5 Vdc, 0-10 Vdc, 0-1 mA (source or sink), and 4-20 mA (source or sink). The top and bottom of each output range can be scaled to fit a user-selected display span.

CONTROL OUTPUT OPTIONS

Additional outputs can be provided by a horizontal upper- board. Available options include single-setpoint control with one 10 A relay, dual-setpoint control with two 10 A relays, 4-20 mA proportional control (source or sink), time-proportional 2-A solid-state relay control, and isolated, parallel BCD output. For additional information, please refer to the QUANTA and Mechanical sections.

SPECIFICATIONS
SCALING , Q2000H
At min scale factor
At max scale factor
Scale factor
0.05 Hz/count
10 Hz/count
Maximum display
1,999 counts
1,999 counts
Frequency for max display
100 Hz
20 kHz
Minimum usable Frequency
5 Hz
1 kHz
Zero suppression
­200 to +40 cts
­200 to +40 cts

SCALING , Q2000H
At min scale factor
At max scale factor
Scale factor
0.01 Hz/count
2Hz/count
Maximum display
9,999 counts
9,999 counts
Frequency for max display
100 Hz
20 kHz
Minimum usable Frequency
5 Hz
1 kHz
Zero suppression
­1000 to +200 cts
­1000 to +200 cts

SIGNAL INPUT
Configuration: Single-ended
Input resistance: 150 kohm
Maximum signal: 130 V rms
Minimum signal, 10 mV hysteresis setting: 15 mV p-p, 100 Hz to 2 kHz FS, 30 mV p-p, 2 kHz to 5 kHz FS 90 mV p-p, 5 kHz to 20 kHz FS
Minimum signal: 100 mV hysteresis setting,125 mV p-p, 100 Hz to 5 kHz FS,175 mV p-p, 5 kHz to 20 kHz FS
Hysteresis: 10 mV or 100 mV
Trigger level: ­1 to +5 V with 100 mV hysteresis or­0.1 to +0.5 V with 10 mV hysteresis
Signal types: TTL logic, CMOS logic, open-collector, 2-wire NPN, 3-wire NPN, 3-wire PNP,2-wire NAMUR, passive magnetic,(see hookup diagrams)

Pulses and  Recoder
A typical application of QUANTA Q2000H AND Q9000H: Display of speed in appropriate engineering units plus relay outputs for alarm and control. The recorder output is standard.

ADJUSTMENTS

Coarse scale factor: Nine ranges selectable via jumper
Fine scale factor: Front-panel accessible potentiometer
Hysteresis & trigger level: Two ranges selectable via jumper or at concertor
Fine trigger level: Front-panel accessible potentiometer
Zero offset: Front-panel accessible potentiometer


NOISE REJECTION
CMR: 120 dB,Dc to 60 Hz
CMV: 1500 Vp per HV test, 354 Vp per IEC spacing

ACCURACY AT 25°C

Error, max: ±0.1% of full-scale frequency ±2 counts
Reading tempco: ±0.01% of reading/°C
Zero tempco: ±0.1 count/°C
Step response: 3 s to 99.9% of span
Warmup to rated accuracy: 10 min (Q2000H), 30 min (Q9000H)

CONVERSION

Technique: Dual-slope, average-value
Signal integration period: 100 ms
Read rate: 2.5/s

DISPLAY
LED: 0.56 in (14.2 mm) 7-segment, red
LCD: 0.50 in (12.7 mm) 7-segment liquid crystal
Symbols: 1.8.8.8 (Q2000H), ­8.8.8.8 (Q9000H)
Decimal-point positions: Three positions selectable by jumpers behind lens or at connector
Overrange indication:
Q2000H: Three least-significant digits blanked
Q9000H, 10,000 - 19,999 counts: Four least-significant digits flash reading Q9000H, over 19,999 counts: All four digits flash zeros

POWER
AC voltages:
120, 240, 100 or 24 Vac +10%/­15%
AC frequency: 49-440 Hz
DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated to 300 Vp; 5 Vdc ±5%, non-isolated
DC power consumption: 5 W max

ENVIRONMENTAL
Operating temperature:
0 to 60°C
Storage temperature: ­40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)

MECHANICAL
Dimensions: Newport DIN1A (1/8 DIN) case (See Mechanical section for drawings)
Weight: 17 oz (480 g)
Case material: 94 V-0 UL-rated polycarbonate



To Order (Specify Model Number, see variations below in partnumber builder table)    
 Part Number   Availability   Price    Description    Qty.
 Q2000-H in-stock    $310.00     Quanta meter,3-1/2 digit display, red LED display, 120 Vac power, 1 mV per count analog output, frequency/rate input (no control output)  
 Q9000-H 1 Week    $350.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 1 mV per count analog output, frequency/rate input (no control output)  
 Q2020-H in-stock    $380.00     Quanta meter,3-1/2 digit display, red LED display, 120 Vac power, 0 to 10 Vdc analog output, frequency/rate input (no control output)  
 Q9001-H in-stock    $515.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 1 mV per count analog output, dual setpoint 10 A relay, frequency/rate input  
 
All dollar amounts on this site are shown in US currency.

 Part Number Builder
(1) (2) (3) (4) (5) (6) (7)
Q -
Option Descriptions
(1) Display Resolution and Case Style
  Select
2 for ±1,999 counts (3 1/2 digits)
3 for ±1,999 with INFINITY FAMILY BEZEL AND ADAPTER
9 for ±9,999 counts (4 digits)
8 for ±9,999 with INFINITY FAMILY BEZEL AND ADAPTER
(2) Display Type & Meter Power - LCD
  Select
0 for LED, 120 Vac (49/440 Hz) (BQ20 or BQ90)
1 for LCD, 120 Vac (49/440 Hz) (BQ21)
2 for LED, 240 Vac (49/440 Hz) (BQ22 or BQ92)
3 for LCD, 240 Vac (49/440 Hz) (BQ23)
4 for LED, 9-32 Vdc (isolated) (BQ24 or BQ94)
5 for LCD, 9-32 Vdc (isolated) (BQ25)
6 for LED, 5 Vdc** (BQ26 or BQ96)
7 for LCD, 5 Vdc** (BQ27)
8 for LED, 24 Vac (BQ28 or BQ98)
9 for LCD, 24 Vac (BQ29)
A for LED, 26-56 Vdc (isolated) (BQ2A or BQ9A)
B for LCD, 26-56 Vdc (isolated) (BQ2B)
(3) Analog Output Adders
  Select
0 for ±1 or ±2 V recorder output, non-isolated. (standard)
1 for 0-5 Vdc, non-isolated (BA01)
2 for 0-10 Vdc, non-isolated (BA01)
3 for 0-1 mA, source or sink, non-isolated (BA01)
4 for 4-20 mA, source or sink, non-isolated (BA01)
5 for 4-20 mA, sink (high-compliance), non-isolated. (BA02)
6 for 4-20 mA, ISOLATED (BA03)
(4) Control Output Adders
  Select
0 for None
1 for Dual-setpoint 10 A relay. (BDT1)
2 for Proportional 4-20 mA control, source or sink (BDP1)
3 for Time-proportional solid-state 2 A relay plus proportional 4-20 mA control, source or sink. (BDP2)
4 for BCD (isolated, latched, buffered, byte addressable). Includes 50-pin mass-termination connector. (BDD2)
5 for Single-setpoint 10 A relay (BDS1)
(5) Signal Conditioning Input Adders
  Select
H for Frequency/rate input - Default calibration 100mV Hystersis 1000 Hz = 1000 (Q2000/Q3000) or 1000 Hz = 0-9999 (Q8000/Q9000) (BSCH1)
(6) Display Color
  Select
Nothing(leave field blank) for Red Display (standard)
,G for Green Display
(7) Lens
  Select
Nothing(leave field blank) for Standard Lens (with Newport logo)
,BL for Blank Lens
 NOTE: All combinations may not be valid, check spec sheet for valid part numbers.


   
Keywords: Frequency Meter, Tachometer & Rate Meter

   Back to top

Sitemap | Order Status | Careers | About Us | ISO 9001 | Worldwide Affiliates | Terms & Conditions
Copyright ©2012, NEWPORT Electronics, Inc. All rights reserved.