The QUANTA 9000S is a 4-digit meter/controller for use with
pressure transducers and load cells which require bridge excitation
and preamplification of microvolt signals. Deadload and tare adjustments
make it ideal for weighing applications. Like process meters, it provides
zero and span adjustments for direct readout in engineering units.
It can also be used as a microvoltmeter with differential input.
EXCITATION
SUPPLY AND PREAMPLIFIER
The Q9000S provides a constant-voltage excitation supply, which is
adjustable from 1 to 10 V and can drive up to 30 mA. It also provides
a high-impedance preamplifier, which is continuously adjustable so
that signal levels from ±2.0 to ±500 mV can produce the
full-scale display of ±9,999 counts. At 10 V excitation and maximum
gain, full-scale readout can be obtained with strain-gauge sensitivity
as low as 0.20 mV/V. An active filter is provided for noise reduction.
BRIDGE CONNECTION
Bridges can be connected by 4 or 6 wires. In 6-wire connection, two
lines sense the voltage applied to the bridge and compensate for lead
resistance (these two sense lines require the optional PCB connector
D1).
RESOLUTION
AND ACCURACY
Resolution is one part in ±9,999 counts, or 0.01% of full scale.
Accuracy is 99.9% of reading. The measurement is ratiometric to reduce
errors due to temperature and power- line variations. The same internal
reference controls the meter gain and the excitation level.
POWER OPTIONS
Six types of meter power are available: 120 Vac, 240 Vac, 24 Vac,
5 Vdc, isolated 9-32 Vdc, or isolated 26-56 Vdc.
|
ANALOG
OUTPUT OPTIONS
A
0.2 mV/count (±2 V full-scale) analog output is standard
and is useful for driving a strip-chart recorder. An additional
analog output can be provided by an optional vertical plug-in
board. Available output signals are 0-5 Vdc, 0-10 Vdc, 0-1
mA (source or sink), and 4-20 mA (source or sink). The top
and bottom of each output range can be scaled to fit a user-selected
display span.
Additional outputs can be provided by a horizontal upper board.
Available options include single-setpoint control with one
10 A relay, dual-setpoint control with two 10 A relays, 4-20
mA proportional control (source or sink), time-proportional
2 A solid-state relay control, and isolated, parallel BCD
output. For additional information, please refer to the QUANTA,
Options, and Mechanical sections.
|
|
SPECIFICATIONS
|
ANALOG INPUT
|
Most-sensitive
scaling
|
Least-sensitive
scaling
|
|
Range
|
±2.0
mV
|
±500
mV
|
|
Resolution
|
0.2
µV/count
|
50
µV/count
|
|
Preamplifier
gain
|
405
|
4.01
|
|
Post-amplifier
gain
|
2.47
|
1.00
|
Gain adjustment:
Seven overlapping ranges with fine adjustment
Input configuration: Differential, ratiometric
Zero adjustment: ±225 µV/volt of excitation
Span adjustment: 1 to 2.47 of preamp gain
Bias current: 11 nA typ, 22 nA max
NOISE REJECTION
NMR, SIG HI to SIG LO: 150 dB, 50/60 Hz
CMR, SIG GND to SIG LO: 80 dB, DC to 60 Hz
CMV, SIG GND to SIG LO: ±1 V, DC to 60 Hz
CMR, SIG GND to PWR GND: 120 dB
CMV, SIG GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC
spacing
ACCURACY AT 25°C
Error, max: ±0.05% of reading ±2 counts
Span tempco: ±0.01% of reading/°C
Zero drift, max: ±0.3 µV/°C
Offset drift, max: ±(0.01% offset V ±0.01% FS V)/°C
Step response: 1 s to 99.9% of span
Warmup to rated accuracy: 60 min
CONVERSION
Technique: Dual-slope, average-value
Signal integration period: 100 ms
Read rate: 2.5/s
EXCITATION
Configuration: Constant-voltage
Voltage sense: Internal or remote
Voltage range: Adjustable from 1 to 10 V
Load current, max: 30 mA
DISPLAY TYPE
Type: 7-segment, red
Height: 0.56 in (14.2 mm)
Symbols: 8.8.8.8
Decimal point positions: Three positions selectable by jumpers
behind lens or at connector
0 to ±9,999 counts: Normal operation
±10,000 to ±19,999 counts: Four least-significant
digits flash reading
Beyond ±19,999 counts: All four digits flash zeros
Overrange indication: Three least-significant digits blanked
POWER
AC voltages: 120, 240 or 24 Vac +10%/15%
AC frequency: 49-440 Hz
DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated
to 300 Vp; 5 Vdc ±5%, non-isolated
Power consumption: 10 W max
ENVIRONMENTAL
Operating temperature: 0 to 60°C
Storage temperature: 40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)
MECHANICAL
Dimensions: Newport DIN1A (1/8 DIN) case (see Mechanical
section for drawings)
Weight: 17 oz (480 g)
Case material: 94V-0 UL-rated polycarbonate
|
|
|
† All dollar amounts on this site are shown in US currency.
|
|
 |
 |
 |
 |
 |
 |
 |
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
Display Resolution and Case Style
|
| |
Select 9 for ±9,999 counts (4 digits) 8 for ±9,999 with INFINITY FAMILY BEZEL AND ADAPTER
|
(2)
|
Display Type & Meter Power - LCD
|
| |
Select 0 for LED, 120 Vac (49/440 Hz) (BQ20 or BQ90) 1 for LCD, 120 Vac (49/440 Hz) (BQ21) 2 for LED, 240 Vac (49/440 Hz) (BQ22 or BQ92) 3 for LCD, 240 Vac (49/440 Hz) (BQ23) 4 for LED, 9-32 Vdc (isolated) (BQ24 or BQ94) 5 for LCD, 9-32 Vdc (isolated) (BQ25) 6 for LED, 5 Vdc** (BQ26 or BQ96) 7 for LCD, 5 Vdc** (BQ27) 8 for LED, 24 Vac (BQ28 or BQ98) 9 for LCD, 24 Vac (BQ29) A for LED, 26-56 Vdc (isolated) (BQ2A or BQ9A) B for LCD, 26-56 Vdc (isolated) (BQ2B)
|
(3)
|
Analog Output Adders
|
| |
Select 0 for ±1 or ±2 V recorder output, non-isolated. (standard) 1 for 0-5 Vdc, non-isolated (BA01) 2 for 0-10 Vdc, non-isolated (BA01) 3 for 0-1 mA, source or sink, non-isolated (BA01) 4 for 4-20 mA, source or sink, non-isolated (BA01) 5 for 4-20 mA, sink (high-compliance), non-isolated. (BA02) 6 for 4-20 mA, ISOLATED (BA03)
|
(4)
|
Control Output Adders
|
| |
Select 0 for None 1 for Dual-setpoint 10 A relay. (BDT1) 2 for Proportional 4-20 mA control, source or sink (BDP1) 3 for Time-proportional solid-state 2 A relay plus proportional 4-20 mA control, source or sink. (BDP2) 4 for BCD (isolated, latched, buffered, byte addressable). Includes 50-pin mass-termination connector. (BDD2) 5 for Single-setpoint 10 A relay (BDS1)
|
(5)
|
Signal Conditioning Input Adders
|
| |
Select S for Strain-gage/microvolt input plus excitation
|
(6)
|
Display Color
|
| |
Select Nothing(leave field blank) for Red Display (standard) ,G for Green Display
|
(7)
|
Lens
|
| |
Select Nothing(leave field blank) for Standard Lens (with Newport logo) ,BL for Blank Lens
|
| NOTE: All combinations may not be valid, check spec sheet for valid part numbers. |
| | | | | |
|
 |
 |
|
|
|
|